Search Results (2 CVEs found)

CVE Vendors Products Updated CVSS v3.1
CVE-2019-25717 1 Draeger 2 Infinity Delta, Infinity Kappa 2026-06-03 4.3 Medium
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.
CVE-2019-25716 1 Draeger 2 Infinity Delta, Infinity Kappa 2026-06-02 6.5 Medium
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.